Thank you! Your submission has been received!
Oops! Something went wrong while submitting the form.

PARAMETERSPECIFICATION
Main BodyHigh-strength engineering plastic rack, light, beautiful and corrosion resistant
Scope of ApplicationAdult, child
Gas SourceO2: 0.27–0.55Mpa;

N2O: 0.27–0.55MPa.

Flow MeterO2: 0.1–1.0L/min; 1.1–10L/min

N20: 0.1–1.0L/min; 1.1–10L/min

Air: 0.1–10L/min

O2, N2O Linkage and N2O

Stopper

When using nitrous oxide, oxygen concentration>25%;

When the oxygen pressure < 0.2KPa, the flow of nitrous oxide would be cut off.

Flow Rate of Rapid Oxygen Supply25–75L/min
Low Oxygen Pressure AlarmThere will be sound alarm when the oxygen pressure < 0.2MPa
VaporizerIt has the function of automatic compensation based on pressure, temperature, and flow rate. The regulation range of evaporator concentration is 0-5vol%.

Among Halothane, Enflurane, Isoflurane and Sevoflurane, one can be chosen for application as required by the customer. And vaporizers that are imported with original packaging are also available.

Respiratory CircuitWorking mode: all-close, semi-close, semi-open;

APL: 0.5-7kPa

Respiratory BellowsBellows for adults, tidal volume range:0–1500ml;

Bellows for children, tidal volume range:0-300ml.

Anyone can be chosen for application as required by the customer

VENTILATOR
Display Mode10.4” Touch Screen, High-definition TFT color LCD
Ventilation ModeIPPV, SIPPV, SIMV, PEEP, MANUAL, SIGH, IMV, VCV, PCV
Ventilation ParametersTidal volume           20–1500ml

Rate                  1–99bpm

SIMV rate                    2–20bpm

I:E                   4:1–1:8

Inspiratory trigger pressure    -1.0–2.0kPa

PEEP                  0–2.0kPa

Pressure range          1.0–6.0kPa

SIGH         1.5 times the inspiratory time, 70-150 adjustable

Inspiratory platform      0–50%

Parameters for Ventilation

Monitoring

Tidal volume, ventilation volume, IPPV rate, SIMV rate, total respiratory rate, I/E, peak pressure of airway, average pressure, pressure – time waveform, flow rate – time waveform, PEEP, inspiratory trigger pressure, Inspiratory platform, autonomous inspiratory rate, lung compliance, airway resistance, flow-volume loop, pressure-volume loop
Monitoring of Oxygen

Concentration

21%–100%
SECURITY ALARM SYSTEM
Oxygen Concentration AlarmUpper limit setting range   21%–100%;

Low limit setting range     10%–80%

Airway Pressure AlarmUpper limit setting range   0.3–6.0 kPa;

Low limit setting range     0.2–5.0 kPa.

Per-minute Ventilation Volume AlarmUpper limit setting range   3.0–30L/min;

Low limit setting range     1.0–10L/min.

Sustained High-pressure

Alarm

It will give alarm when stress have consistently been higher than 2.5 kPa
Suffocation AlarmIt will give sound and light alarm if there is no tidal volume input for 15 seconds
POWER ALARM
PowerAC 220V±10%, 50Hz±1Hz, Storage battery (UPS)
Mechanical Arms (End-expiratory CO2 monitoring)
Note: Items within brackets are optional.

ARI-800C Patient Monitor

12 inch Six Standard parameters: ECG, RESP, NIBP, SPO2, 2_TEMP, PR/HR