Technical Specifications

PARAMETERSPECIFICATION
Main BodyHigh-strength engineering plastic rack, light, beautiful and corrosion resistant
Scope of ApplicationAdult, child
Gas SourceO2: 0.27–0.55Mpa;

N2O: 0.27–0.55MPa.

Flow MeterO2: 0.1–10L/min;

N2O: 0.1–10L/min

O2, N2O Linkage and N2O

Stopper

When using nitrous oxide, oxygen concentration>25%;

When the oxygen pressure < 0.2MPa, the flow of nitrous oxide would be cut off.

Flow Rate of Rapid Oxygen Supply25–75L/min
Low Oxygen Pressure AlarmThere will be sound alarm when the oxygen pressure < 0.2MPa
VaporizerIt has the function of automatic compensation based on pressure, temperature, and flow rate. The regulation range of evaporator concentration is 0-5vol%.

Among Halothane, Enflurane, Isoflurane and Sevoflurane, one can be chosen for application as required by the customer. And vaporizers that are imported with original packaging are also available.

Respiratory CircuitWorking mode: all-close, semi-close, semi-open;

Releasing pressure: 0.5-7kPa.

Respiratory BellowsBellows for adults, bellows for children, tidal volume range 0-1500ml
VENTILATOR
Display ModeHigh-definition 5.7” LCD screen display
Ventilation ModeIPPV, SIPPV, SIMV, PEE, MANUAL, SIGH
Ventilation ParametersTidal volume           50–1500ml

Rate                  2–99bpm

SIMV rate                   2–20bpm

I:E                  2:1–1:8

Inspiratory trigger pressure   -1.0–2.0kPa

PEEP                 0–2.0kPa

Pressure range         1.0–6.0kPa

SIGH                 1.5 times the inspiratory time

Parameters for Ventilation

Monitoring

Tidal volume, ventilation volume, IPPV rate, SIMV rate, total respiratory rate, I/E, peak pressure of airway, pressure – time waveform, flow rate – time waveform, PEEP, inspiratory trigger pressure.
Monitoring of Oxygen

Concentration

(21%–100%)
SECURITY ALARM SYSTEM
Oxygen Concentration alarmUpper limit setting range    (50%–100%)

Low limit setting range      (15%–50%)

Airway pressure alarmUpper limit setting range   1.0–6.0kPa;

Low limit setting range     0.4–2.0kPa.

Per-minute ventilation volume alarmUpper limit setting range   3.0–30L/min;

Low limit setting range     1.0–10L/min.

Sustained high-pressure

alarm

It will give alarm when stress have consistently been higher than 2.5 kPa
Suffocation alarmIt will give sound and light alarm if there is no tidal volume input for 15 seconds
POWER ALARM
PowerAC 220V±10%, 50Hz±1Hz, (Storage battery, UPS)
Note: Items within brackets are optional.