Thank you! Your submission has been received!
Oops! Something went wrong while submitting the form.

PARAMETERSPECIFICATION
Main BodyHigh-strength engineering plastic rack, light, beautiful and corrosion resistant
Scope of ApplicationAdult, child
Gas SourceO2 0.27–0.55MPa
Flow MeterO2: 0.1–1.0L/min; 1.1–10L/min
Flow Rate of Rapid Supply25–75L/min
Low Oxygen Pressure AlarmThere will be sound alarm when the oxygen pressure < 0.2MPa
VaporizerIt has the function of automatic compensation based on pressure, temperature, and flow rate. The regulation range of evaporator concentration is 0-5vol%.

Among Halothane, Enflurane, Isoflurane and Sevoflurane, one can be chosen for application as required by the customer.

Respiratory CircuitWorking mode: all-close, semi-close, semi-open

Releasing pressure: 0.5-7kPa.

Respiratory BellowsBellows for adults, bellows for children, tidal volume range:0-1500ml
VENTILATOR
Display ModeHigh-definition 5.7” LCD screen display
Ventilation ModeIPPV, MANUAL (Manual monitoring of tidal, volume, ventilation volume, respiratory rate).
Ventilation ParametersTidal volume           50–1500ml

Rate                  2–99bpm

SIMV rate                   2–20bpm

I:E                  2:1–1:8

Inspiratory trigger pressure   -1.0–2.0kPa

Pressure range        1.0–6.0kPa

Parameters for Ventilation

Monitoring

Tidal volume, ventilation volume, IPPV rate, I/E, peak pressure of airway, inspiratory trigger pressure, pressure – time waveform.
SECURITY ALARM SYSTEM
Airway Pressure AlarmUpper limit setting range   1.0–6.0kPa;

Low limit setting range     0.4–2.0kPa.

Per-minute Ventilation Volume AlarmUpper limit setting range   3.0–30L/min;

Low limit setting range     1.0–10L/min.

Sustained High-pressure

Alarm

It will give alarm when stress have consistently been higher than 2.5 kPa
Suffocation AlarmIt will give sound and light alarm if there is no tidal volume input in 15s
POWER ALARM
PowerAC 220V±10%   50Hz±1Hz